FESSH prep Imaging Radiographs and fluoroscopy

Radiation safety with intraoperative fluoroscopy in hand surgery

Last reviewed v1

FESSH-161Imaging●○○ basic

Regarding radiation exposure during intraoperative fluoroscopy in hand and wrist surgery:

  1. 1.Using a mini C-arm instead of a standard C-arm reliably lowers the radiation dose to the surgeon's hands.

  2. 2.With a horizontal beam, a surgeon standing on the image-intensifier side receives less scattered radiation than one standing on the X-ray tube side.

  3. 3.Placing the hand directly on the image intensifier, rather than raising it towards the X-ray tube, lowers radiation exposure to both the patient and the surgeon.

  4. 4.Doubling the distance from the radiation source reduces exposure to about one half.

  5. 5.Radiation-attenuating gloves make it safe to hold the hand in the primary beam.

FESSH marking: +1 per correct statement, 0 for blank, −1 for incorrect. Click the same TRUE/FALSE button twice to clear it.

Question ID
FESSH-161
First published
Last reviewed
Version
v1
Answer key and explanations — open after answering
  1. 1. Using a mini C-arm instead of a standard C-arm reliably lowers the radiation dose to the surgeon's hands.

    False

    Device type alone does not determine hand dose; geometry and hand position do. Kaplan and colleagues' review notes that the mini C-arm reduces overall exposure to the surgeon in some studies but may increase the dose to the hands, which are often in or near the direct beam. In Vosbikian and colleagues' practical comparison of two surgeons' case series, ring-dosimeter absorption to the hand was about double with the mini C-arm, and under identical vertical geometry Singer and colleagues measured 53% to 70% more direct exposure with the mini C-arm, which has the shorter source-to-intensifier distance. Athwal and colleagues, by contrast, found lower exposure with the mini C-arm in the clinical configurations they tested, so the comparison depends on the set-up.

    • Kaplan DJ, Patel JN, Liporace FA, Yoon RS. Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg. 2016;10:27. PMID 27999617
    • Vosbikian MM, Ilyas AM, Watson DD, Leinberry CF. Radiation exposure to hand surgeons' hands: a practical comparison of large and mini C-arm fluoroscopy. J Hand Surg Am. 2014;39(9):1805-9. PMID 25086796
    • Singer G, Herron B, Herron D. Exposure from the large C-arm versus the mini C-arm using hand/wrist and elbow phantoms. J Hand Surg Am. 2011;36(4):628-31. PMID 21414730
    • Athwal GS, Bueno RA, Wolfe SW. Radiation exposure in hand surgery: mini versus standard C-arm. J Hand Surg Am. 2005;30(6):1310-6. PMID 16344194
  2. 2. With a horizontal beam, a surgeon standing on the image-intensifier side receives less scattered radiation than one standing on the X-ray tube side.

    True

    Scatter is greatest between the X-ray tube and the patient, so for horizontal-beam (lateral) imaging, standing on the side opposite the tube greatly reduces scatter exposure (Kaplan and colleagues). In a simulated wrist-surgery set-up with a cadaveric upper limb, Athwal and colleagues likewise found that the safest mini C-arm configuration was with the surgeon on the image-intensifier side rather than the source side, and advised standing behind the lead-encased intensifier.

    • Kaplan DJ, Patel JN, Liporace FA, Yoon RS. Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg. 2016;10:27. PMID 27999617
    • Athwal GS, Bueno RA, Wolfe SW. Radiation exposure in hand surgery: mini versus standard C-arm. J Hand Surg Am. 2005;30(6):1310-6. PMID 16344194
  3. 3. Placing the hand directly on the image intensifier, rather than raising it towards the X-ray tube, lowers radiation exposure to both the patient and the surgeon.

    True

    The image intensifier should be as close to the patient as possible, which keeps the limb away from the tube (Kaplan and colleagues). With the intensifier below cadaver hand-wrist and elbow specimens, Singer and colleagues found that raising the specimen on a hand table increased exposure by 80% to 94% compared with placing it directly on the intensifier, and concluded that direct exposure to the patient and scatter to the surgeon are minimised with the extremity on the intensifier. Imaging a cadaver ankle, Giordano and colleagues also found patient and surgeon exposure notably amplified when the specimen was closer to the source.

    • Kaplan DJ, Patel JN, Liporace FA, Yoon RS. Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg. 2016;10:27. PMID 27999617
    • Singer G, Herron B, Herron D. Exposure from the large C-arm versus the mini C-arm using hand/wrist and elbow phantoms. J Hand Surg Am. 2011;36(4):628-31. PMID 21414730
    • Giordano BD, Baumhauer JF, Morgan TL, Rechtine GR. Patient and surgeon radiation exposure: comparison of standard and mini-C-arm fluoroscopy. J Bone Joint Surg Am. 2009;91(2):297-304. PMID 19181973
  4. 4. Doubling the distance from the radiation source reduces exposure to about one half.

    False

    Radiation intensity follows the inverse-square law: doubling the distance from the source reduces exposure to about one quarter, not one half (Kaplan and colleagues). Stepping back from the beam is therefore one of the simplest ways for the surgical team to reduce scatter dose, alongside shorter screening time, pulsed rather than continuous fluoroscopy, and collimation of the beam.

    • Kaplan DJ, Patel JN, Liporace FA, Yoon RS. Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg. 2016;10:27. PMID 27999617
  5. 5. Radiation-attenuating gloves make it safe to hold the hand in the primary beam.

    False

    The hands receive the greatest direct exposure and are the hardest part to protect. Kaplan and colleagues' review notes that attenuating gloves give much less protection than aprons or thyroid shields, that gloves in the beam can increase scatter and exposure to the hand inside them, and that gloves cannot substitute for keeping the hands out of the beam, for example by holding the limb with an instrument. In a phantom study with standardised mini C-arm screening, Cantlon and Ilyas found that gloves and attenuating cream lowered index-finger dose compared with a standard surgical glove, and recommended them in addition to other exposure-reduction measures. In a mannequin model of mini C-arm use, Chapman and colleagues found that the hand received far more radiation than any other measured site.

    • Kaplan DJ, Patel JN, Liporace FA, Yoon RS. Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg. 2016;10:27. PMID 27999617
    • Cantlon MB, Ilyas AM. Assessment of Radiation Protection in Hand-Shielding Products With Mini C-Arm Fluoroscopy. Hand (N Y). 2021;16(4):505-510. PMID 31402748
    • Chapman T, Martin DP, Williamson C, Tinsley B, Wang ML, Ilyas AM. Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?. Hand (N Y). 2018;13(5):552-557. PMID 28643518