FESSH prep Wrist trauma Distal radius fractures
Tendon complications of volar locking plate fixation
Regarding flexor and extensor tendon complications after volar locking plate fixation of distal radius fractures:
1.On Soong's lateral-radiograph grading, a grade 2 plate lies directly on, or distal to, the volar rim of the distal radius.
2.When practising surgeons grade lateral radiographs, Soong's classification of volar plate prominence shows high interobserver agreement.
3.Extensor pollicis longus is the extensor tendon most often ruptured after volar locking plate fixation.
4.Dorsal cortical penetration by distal screws is detected more reliably on the standard lateral fluoroscopic view than on additional views such as the dorsal tangential view.
5.Removal of a volar plate is advised when it impinges at the watershed line or causes flexor tenosynovitis.
FESSH marking: +1 per correct statement, 0 for blank, −1 for incorrect. Click the same TRUE/FALSE button twice to clear it.
Answer key and explanations — open after answering
-
1. On Soong's lateral-radiograph grading, a grade 2 plate lies directly on, or distal to, the volar rim of the distal radius.
True
Soong and colleagues drew a line tangential to the most volar extent of the volar rim, parallel to the volar cortex of the shaft: grade 0 plates do not extend volar to this line, grade 1 plates lie volar to it but proximal to the rim, and grade 2 plates lie directly on or distal to the rim. Two of their three flexor tendon ruptures occurred with grade 2 prominence, at the watershed line where the flexor tendons lie closest to the plate.
- Soong M, Earp BE, Bishop G, Leung A, Blazar P. Volar locking plate implant prominence and flexor tendon rupture. J Bone Joint Surg Am. 2011;93(4):328-35. PMID 21239658
-
2. When practising surgeons grade lateral radiographs, Soong's classification of volar plate prominence shows high interobserver agreement.
False
When 115 international fracture and upper-limb surgeons graded lateral radiographs in Crijns and colleagues' study, agreement on the Soong grade was only fair (kappa 0.32), and agreement on whether the plate was more prominent than the watershed line was lower still. Separation of the plate from the bone and dorsal angulation of more than 5° were graded more reliably.
- Crijns TJ, Merkel P, Kortlever JTP, Wagner KJ, Ring D, Vagner G, et al.. The Interobserver Reliability of Grading of Distal Radius Volar Plate Prominence. J Hand Surg Am. 2022;47(11):1095-1100. PMID 36075822
-
3. Extensor pollicis longus is the extensor tendon most often ruptured after volar locking plate fixation.
True
In Kunes and colleagues' systematic review of extensor tendon injury after volar locking plate fixation, extensor pollicis longus was the tendon most commonly ruptured, and dorsal screw prominence and fracture fragments were among the risk factors; Azzi and colleagues' pooled data put extensor pollicis longus rupture at 0.8% after volar plating. In Ohno and colleagues' prospective series, distal screws were chosen 2 mm shorter than measured, yet some screws still penetrated the dorsal cortex and two extensor pollicis longus ruptures occurred without screw penetration, so shorter screws do not abolish the risk. Cadaver testing by Baumbach and colleagues suggests that distal screws at 75% of the measured length keep similar primary stability, which gives a biomechanical basis for choosing shorter screws.
- Kunes JA, Hong DY, Hellwinkel JE, Tedesco LJ, Strauch RJ. Extensor Tendon Injury After Volar Locking Plating for Distal Radius Fractures: A Systematic Review. Hand (N Y). 2022;17(1_suppl):87S-94S. PMID 35168382
- Azzi AJ, Aldekhayel S, Boehm KS, Zadeh T. Tendon Rupture and Tenosynovitis following Internal Fixation of Distal Radius Fractures: A Systematic Review. Plast Reconstr Surg. 2017;139(3):717e-724e. PMID 28234845
- Ohno K, Takigawa N, Watanabe C, Oda A, Ishizu T, Neo M. Effect of Downsized Screw Selection on Bone Healing and Postoperative Complications in Volar Plating of Distal Radius Fractures. Orthopedics. 2021;44(2):e259-e265. PMID 33316819
- Baumbach SF, Synek A, Traxler H, Mutschler W, Pahr D, Chevalier Y. The influence of distal screw length on the primary stability of volar plate osteosynthesis--a biomechanical study. J Orthop Surg Res. 2015;10:139. PMID 26351239
-
4. Dorsal cortical penetration by distal screws is detected more reliably on the standard lateral fluoroscopic view than on additional views such as the dorsal tangential view.
False
Bergsma and colleagues' systematic review found that additional fluoroscopic views detect dorsal screw penetration better than conventional views, and that the dorsal tangential view is the most studied of them and is practical and time-efficient. In Hill and colleagues' cadaver study, surgeons detected penetration with a sensitivity of 58% on the lateral view, against 67% on the dorsal tangential view and 88% on a 45° supination view. Even so, in Ganesh and colleagues' clinical series CT found further prominent screws missed on the dorsal tangential view, so they advise choosing distal screws 2–4 mm shorter than measured.
- Bergsma M, Denk K, Doornberg JN, van den Bekerom MPJ, Kerkhoffs GMMJ, Jaarsma RL, et al.. Volar Plating: Imaging Modalities for the Detection of Screw Penetration. J Wrist Surg. 2019;8(6):520-530. PMID 31815069
- Hill BW, Shakir I, Cannada LK. Dorsal Screw Penetration With the Use of Volar Plating of Distal Radius Fractures: How Can You Best Detect?. J Orthop Trauma. 2015;29(10):e408-13. PMID 26402306
- Ganesh D, Service B, Zirgibel B, Koval K. The Detection of Prominent Hardware in Volar Locked Plating of Distal Radius Fractures: Intraoperative Fluoroscopy Versus Computed Tomography. J Orthop Trauma. 2016;30(11):618-621. PMID 27769074
-
5. Removal of a volar plate is advised when it impinges at the watershed line or causes flexor tenosynovitis.
True
Floquet and colleagues' systematic review of 145 patients with flexor tendon rupture after volar plating found the cause usually mechanical, from the plate or protruding screws, with plate impingement related to placement beyond the watershed line; flexor pollicis longus was the tendon most often injured, at a median of 8 months after surgery. They recommend placing the plate proximal to the watershed line when possible and removing it 4 months after surgery if it impinges by Soong's criteria or if tenosynovitis appears. Soong and colleagues similarly advised avoiding implant prominence at the watershed line, where the flexor tendons lie closest to the plate.
- Floquet A, Druart T, Lavantes P, Vendeuvre T, Delaveau A. Flexor tendon rupture after volar plating of distal radius fracture: A systematic review of the literature. Hand Surg Rehabil. 2021;40(5):535-546. PMID 34033928
- Soong M, Earp BE, Bishop G, Leung A, Blazar P. Volar locking plate implant prominence and flexor tendon rupture. J Bone Joint Surg Am. 2011;93(4):328-35. PMID 21239658